Pogo Automatic - Important Cartridge Handling and Information; Proper Cartridge Storage and Expiration

Pogo Automatic
88 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Open expiration date
(X = the # of days from the time the cartridge foil cup was
opened. The meter will display this as a number.)
Xd
GTIN: 00856176008018
S
i
n
g
l
e
P
a
t
i
e
n
t
U
s
e
O
n
l
y
D
I
S
C
A
R
D
D
A
T
E
w
w
w
.
p
r
e
s
s
p
o
g
o
.
c
o
m
A
R
T
-
0
2
9
4
5
-
B
D
I
S
C
A
R
D
1
5
D
A
Y
S
A
F
T
E
R
O
P
E
N
I
N
G
19
Customer Support 1-855-IMI-POGO (464-7646) • www.presspogo.com
Important POGO Automatic
®
Test Cartridge Information
Please note: Proper storage and handling of test cartridges is
essential to obtain accurate results.
Store test cartridges in a cool, dry place between
40°F – 86°F, and in a relative humidity range of 10% – 90%
(non-condensing).
Ensure meter and cartridge are both at room temperature before
testing (see Meter Specications on page 78).
Keep out of direct sunlight and heat. Do not refrigerate or freeze.
Do not use the test cartridge past the expiration date
( Use by:) printed on cartridge label.
Immediately mark the discard date in the box on cartridge top
(see Example A). Do not use aer the discard date.
Example A: 15-day Discard Date
Opened Expires
April 16 May 1
When you insert a new cartridge, the following
symbol will appear which means you have the # of
days shown on the meter display to use all tests:
Note: 15-day discard date is an example only.
Check label on cartridge top for accurate open
expiration date.