Control Solution Testing
Performing a Control Solution Test helps
you check that the meter and test cartridge
are working properly together, and that your
blood glucose results are reliable.
The POGO Automatic Control Solutions
contain a known amount of glucose. You
perform a test with POGO Automatic
Control Solutions similarly to a blood
glucose test, except that you use Control
Solution instead of blood (see how to
perform a Control Solution Test on
Page 26).
The Control Solution Test result appears
on the meter as a pass or fail result. If the
Control Solution result fails, either you may
not be performing the test correctly or the
meter and test cartridge are not working
properly together.
Either Level 1 or Level 2 Control Solution
can be used to check the system. Obtaining
a PASS result using either is acceptable to
verify system performance. Most users will
nd it sucient to use Level 2.
Applicator
Handle
POGO Automatic
Level 1 Control Solution
Applicator
POGO Automatic
Level 2 Control Solution
Applicator
Control Solution
Tip
NOTE: Control Solution applicators are single-use only.
Use immediately aer opening foil pouch. Discard aer use.