Test Opener
Barcode
Reader
Cartridge Tower
68
Customer Support 1-855-IMI-POGO (464-7646) • www.presspogo.com
Be sure to disinfect all
exterior meter surfaces,
including buttons and
grooves
Cartridge Area
Disinfecting Meter Exterior AFTER
EVERY CARTRIDGE or WEEKLY
(whichever comes rst)
Immediately following cleaning, use a
NEW Clorox Germicidal Wipe to disinfect
ALL external meter surfaces.
Pay special attention to thoroughly wipe
gaps, grooves, and around buttons in
meter case.
Allow the exterior surface to remain
damp for one minute, then let air dry.
Disinfecting Meter Interior
Aer testing, you should check inside of
your meter for blood, including cartridge
door. Clean and disinfect IMMEDIATELY
if blood is present. Any interior part you
cleaned needs to be disinfected, using a
NEW Clorox Germicidal Wipe. Be sure to
disinfect AFTER
EVERY
CARTRIDGE
OR
WEEKLY, whichever comes rst.
(To clean, see Cleaning Meter on page 66).
CAUTION: Do not substitute other household cleaners
(e.g. Windex
®
, Lysol
®
or alcohol) for Clorox
®
Germicidal Wipes
®
to ensure
proper disinfection and to avoid meter damage or malfunction. Other
products, such as Clorox
®
Disinfecting
®
Wipes, have not been tested and
should not be used.