64
Customer Support 1-855-IMI-POGO (464-7646) • www.presspogo.com
Cleaning Test Platform DAILY
• To remove, open the cartridge door by
pressing the button located on the lower
right side of meter.
• Li up and pull from the end closest to
the cartridge door.
• Using a Clorox Germicidal Wipe, clean
both front and back of test port,
including INSIDE test port hole.
Disinfecting Test Platform DAILY
Always make sure any visible blood
and debris is cleaned o test platform
before handling.
(See Cleaning Test Platform DAILY above).
• Using a NEW Clorox Germicidal Wipe,
thoroughly wipe front and back surfaces.
Disinfect the test port hole by pushing
wipe through.
• Allow the test platform surface to remain
damp for one minute, then let air dry
completely.
• Leave o until meter interior is
disinfected.
Test Port Hole
Test Port
Test Platform
Disinfect both front and
back surfaces of test
platform
REMEMBER: Disinfect
inside test port hole
CARING
FOR
YOUR
POGO
AUTOMATIC