25
Customer Support 1-855-IMI-POGO (464-7646) • www.presspogo.com
When Should You Perform
a Control Solution Test?
To practice testing using your POGO Automatic System.
When you open a new tube of POGO Automatic Test Cartridges.
When you think your meter and test cartridges are not working
properly together.
If you have dropped or damaged the meter.
If your cartridges were stored improperly such as in extreme
temperatures or humidity.
If you have been advised by your physician to use
Control Solution.
NOTE: Control Solution Test results are not included in your glucose
averages.
IMPORTANT: When your test results are not consistent with how you feel,
or when you have repeated unexpected results, perform a POGO Automatic
Control Solution test to conrm that your system is working properly.
WARNING: Do not swallow Control Solution; it is not for human
consumption. Do not apply Control Solution to the skin or eyes as it may
cause irritation. The POGO Automatic Control Applicator is a choking
hazard.