Customer Support 1-855-IMI-POGO (464-7646) • www.presspogo.com
Disinfect all
interior meter surfaces,
including cartridge door
Disinfecting Interior
Cartridge Area
Open the cartridge door and remove
cartridge if necessary.
With clockwise motion, wipe all
interior meter walls and all corners
and surfaces of cartridge door with
new Clorox Germicidal Wipe.
It is safe to thoroughly disinfect
all internal surfaces of cartridge
area, including test opener and
barcode reader.
Allow internal surface to remain damp
for one minute, let meter interior air dry.
Meter Optics Care
The meter optics are similar to the
lens of a pair of eyeglasses and may
occasionally collect dust or debris.
Gently wipe optics lens to disinfect.
Be careful not to touch the lens with
your ngers.
Allow optics surface to remain damp
for one minute, then let air dry. Gently
wipe optic lens with a so lint-free
cloth to remove any lm le from
disinfectant wipe.
Gently wipe meter optics
to remove any lm le from
disinfecting
Do not touch lens
directly with ngers
IMPORTANT: Do not apply too much liquid or allow it to run in or around
internal components.