Pass
If the Control Solution Test result is in
range, the screen displays a pass result,
indicating the meter and test cartridge are
working properly together. You are now
ready to perform a blood glucose test.
Fail
If you receive a fail Control Solution Test
result, this indicates that the value is
outside of the Control Solution target
range, and you should immediately repeat
the test.
A failed Control Solution Test may indicate:
• An error in your testing method—
for example, removing applicator from
the test port before indicated.
• Expired or improperly stored Control
Solution, exposure to extreme
temperatures or humidity.
• Expired, damaged or improperly stored
test cartridges.
• A problem with the meter.
WARNING: If your results continue to fail, call Intuity Medical Customer
Support immediately at 1-855-IMI-POGO (464-7646). DO NOT use the
meter to test your blood glucose until you obtain a passing result