The POGO Automatic Meter should only be used with POGO
Automatic Test Cartridges and POGO Automatic Control
Solutions.
Individual test cells and cartridges are single-use only.
Do not attempt to reuse.
Use only fresh whole capillary blood. Do not attempt to use
plasma or serum to obtain an accurate blood glucose test result.
Not to be used for patients who are dehydrated, hypotensive, in
shock, critically ill or in a hyperosmolar state.
The hematocrit range for the POGO Automatic System is
20% – 60%. Extremely high hematocrit levels above 60% may
cause falsely low results and very low hematocrit levels below
20% may cause falsely high results.
The POGO Automatic System has not been evaluated at
altitudes above 10,000 feet.
Additional Information
POGO Automatic Control Solutions should be used for patient
training and educational purposes, and to check that the meter
and test cartridges are working properly.
If the meter is being operated by a second person who is
providing testing assistance to the user, the meter should be
cleaned and disinfected prior to handling by the second person.