MP3
Therapy Beam Analyzer
Operating Manual – Error Messages and Troubleshooting
D175.131.00/11 en 39
PC Software Error Messages
6. All Points are out of Limits Refer to troubleshooting tip 5.
7. An error occurred starting
measurement.
Verify whether the serial PC connection cable is correctly connected.
Verify whether the correct COM port is configured in the PC software.
Verify whether the correct Baud rate is configured in the PC software. I
Is the TBA CONTROL UNIT turned on?
TPR
8. Water is not added or
removed.
Check whether the connection cable to the water reservoir (T43164/K300)
is connected.
Check whether the liquid level switch is correctly connected.
Check whether the water reservoir is connected to mains and the water
pump can be operated via the control buttons at the water tank.
Is the control cable T43164/K1 properly connected between TBA control
unit and junction box?
9. TPR does not stop pump-
ing -> liquid level switch
connected?
Check whether the liquid level switch is connected.
Is the liquid level switch correctly mounted at the holder?