Service Manual – Cleaning
D175.131.00/11 en 45
Service Manual
WARNING
Improper execution of service tasks.
Hazards to Persons! Equipment Damage!
The following service tasks must only be per-
formed by specially trained staff.
Further service tasks may only be carried out by
PTW-Freiburg or by persons authorized by
PTW-Freiburg.
8 Cleaning
8.1 Device Surfaces
First of all, disconnect the MP3 system from the
power line, before attempting to clean the device
surfaces.
Rub the devices down with a dry or slightly moist
cloth. For this purpose use soap water. Do not let
liquid enter the devices. Do not apply spray to the
devices. The devices cannot be sterilized or disin-
fected.
WARNING
Liquids have entered the device.
Shock Hazard, Erroneous Measurement!
The device must be thoroughly inspected by
service personnel before being used again.
CAUTION
Disinfection of the surface with disinfectants on a
phenol base or peroxide compounds.
Equipment Damage!
Do not use disinfectants on a phenol base or
peroxide compounds to disinfect the surface.