Service & Calibration 150821/A4 Page 107 of 110
Measurement Procedure
The 1730T should be set to the following:
[SYSTEM SETUP] [SYSTEM CONFIG]
Integration Cycle = 4
Average = 1
Cable Length = 1M
Const Rsource = 25ohms
[MEAS DISPLAY]
Range: = Auto
Level: 1Volt
Speed: = Slow
Trigger = Manual
BiasV = Off
Binning = Off
Compare = Off
Connections to all standards should be via a 1-meter cable.
A multiple frequency open and short correct should be performed prior to any measurements.
See paragraph 2.5.5 for information on performing an open and short correction.
Connect each standard listed in the Verification Data Sheet and select the test frequency in the
[MEAS DISPLAY] menu of the 1730T. Record the results in the Verification Data Sheet.