EasyManua.ls Logo

QuadTech 1730T - Operation; Bias Current Control

Default Icon
110 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation 150821/A4 Page 49 of 110
BIAS I: 000.000A
When BIAS I is ON, current can be supplied from an external source (1320). To enable/disable
the bias current function, in MEAS DISPLAY (Pg 4/4) press [F2] so that the 000.000 A box is
highlighted. Use the up arrow [] or down arrow [] keys or [F2] to increase/decrease the bias
current function ON/OFF. The instrument default setting is 000.000A (no 1320 connected).
NEXT PAGE 4/4
Ls : 1.2345 H
Q : 1.2345
MEAS DISPLAY
BIAS I:
BIAS I: 000.000 A
ON
F2
Select Bias Current Level
Move highlighted cursor to right
NEXT PAGE 4/4
Ls : 1.2345 H
Q : 1.2345
MEAS DISPLAY
BIAS I:
BIAS I:
ON
000.000 A
NEXT PAGE 4/4
Ls : 1.2345 H
Q : 1.2345
MEAS DISPLAY
BIAS I:
BIAS I:
ON
02
0.000 A
Increase Current Level to 20A
NOTE:
Before BIAS I can be turned ON and a current level set, [LINK 1320] must be set to ON in the
[System Config] menu under [System Setup].
NOTE:
There are no IEEE commands for the BIAS I function on the 1730T instrument. Control of the
bias current is handled via IEEE commands to the 1320 instrument.

Table of Contents