Operation 150821/A4 Page 49 of 110
BIAS I: 000.000A
When BIAS I is ON, current can be supplied from an external source (1320). To enable/disable
the bias current function, in MEAS DISPLAY (Pg 4/4) press [F2] so that the 000.000 A box is
highlighted. Use the up arrow [] or down arrow [] keys or [F2] to increase/decrease the bias
current function ON/OFF. The instrument default setting is 000.000A (no 1320 connected).
NEXT PAGE 4/4
Ls : 1.2345 H
Q : 1.2345
MEAS DISPLAY
BIAS I:
BIAS I: 000.000 A
ON
F2
Select Bias Current Level
Move highlighted cursor to right
NEXT PAGE 4/4
Ls : 1.2345 H
Q : 1.2345
MEAS DISPLAY
BIAS I:
BIAS I:
ON
000.000 A
NEXT PAGE 4/4
Ls : 1.2345 H
Q : 1.2345
MEAS DISPLAY
BIAS I:
BIAS I:
ON
02
0.000 A
Increase Current Level to 20A
NOTE:
Before BIAS I can be turned ON and a current level set, [LINK 1320] must be set to ON in the
[System Config] menu under [System Setup].
NOTE:
There are no IEEE commands for the BIAS I function on the 1730T instrument. Control of the
bias current is handled via IEEE commands to the 1320 instrument.