Interface 150821/A4 Page 85 of 110
3.1.5.22 FORMat[:DATA]
Instruction: FORMat[:DATA]
Parameter: {ASCII}
Return Value: ASK
Function: Set or query the format of the numerical data being transferred.
3.1.5.23 INITiate[:IMMediate]
Instruction: INITiate[:IMMediate]
Parameter: None
Return Value: None
Function: Immediately initiate all sequences.
3.1.5.24 INITiate:CONTinuous
Instruction: INITiate:CONTinuous
Parameter: {ON (1)⏐OFF (0)}
Return Value: {1⏐0}
Function: Set or query if the Trigger is ON or OFF.
Description: ON (1) Turn on Trigger
OFF (0) Turn off Trigger
3.1.5.25 [SENSe]:AVERage:COUNt
Instruction: [SENSe]:AVERage:COUNt
Parameter: {1 ~ 256}
Return Value: {1 ~ 256}
Function: Set or query the averaging number.
3.1.5.26 [:SENSe]:CORRection:COLLect[:ACQuire]:STANdard
Instruction: [:SENSe]:CORRection:COLLect[:ACQuire]:STANdard
Parameter: {1⏐2}
Return Value: None
Function: Perform OPEN or SHORT correction procedure.
Description: 1 Perform OPEN correction procedure
2 Perform SHORT correction procedure
Open Example: SENSe:CORR:COLL:STAN1
3.1.5.27 [:SENSe]:CORRection:COLLect:METHod
Instruction: [:SENSe]:CORRection:COLLect:METHod
Parameter: {REFL2}
Return Value: {REFL2}
Function: Set or query the measured tolerance’s correction procedure.
Description: REFL2 OPEN/SHORT correction procedure