Page 86 of 110 150821/A4 Interface
3.1.5.28 [:SENSe]:CORRection:DATA?STANdard
Instruction: [:SENSe]:CORRection:DATA?STANdard
Parameter: {1⏐2}
Return Value: Two numerical correction values in <NR3> format
Function: Query the numerical correction value.
Description: STANdard 1 OPEN correction value: G, B
STANdard 2 SHORT correction value: R, X
3.1.5.29 [:SENSe]:FIMPedance:APERature
Instruction: [:SENSe]:FIMPedance:APERature
Parameter: 0.025 (Fast), 0.065 (Medium), 0.500 (Slow)
Return Value: 0.025 (Fast), 0.065 (Medium), 0.500 (Slow)
Function: Set or query the measurement speed.
3.1.5.30 [:SENSe]:FIMPedance:RANGe:AUTO
Instruction: [:SENSe]:FIMPedance:RANGe:AUTO
Parameter: {ON (1)⏐OFF (0)}
Return Value: {1⏐0}
Function: Set or query if the Auto Range is ON or OFF.
Description: ON (1) Turn on Auto Range
OFF (0) Turn off Auto Range
3.1.5.31 [:SENSe]:FIMPedance:RANGe[:UPPer]
Instruction: [:SENSe]:FIMPedance:RANGe[:UPPer]
Parameter: {Measured value range⏐UP⏐DOWN}
Unit: [MOHM⏐kOHM⏐OHM⏐mOHM], default is OHM
Return Value: {Measured value range} in <NR3> format
Function: Set or query which Measurement Range is enabled.
Description: UP Go to previous range
DOWN Go to next range
Range Values: 1MΩ, 300kΩ, 100kΩ, 30kΩ, 10kΩ, 3kΩ, 1kΩ, 300Ω, 100Ω, 10Ω, 1Ω, 0.1Ω.
Note: When CONST Rsource is OFF, 300kΩ, 30kΩ, 3kΩ and 300Ω are disabled.
When Test Frequency is set to 100Hz or 120Hz, 0.1Ω is disabled.