Interface Page 63 of 76
3.2.5.13 DISP: WVT, L CT, null, main, system
Instruction: DISP: WVT
DISP: LCT
DISP: NULL
DISP: MAIN
DISP: SYSTem
Parameter: None
Return Value: None
Function: Set the Display to WV Mode, L CT mode, null, main or system mode.
Description: Set the Display to WV Mode
3.2.5.14 LCT:SOUR:VOLT
Instruction: LCT:SOUR:VOLT
Parameter: {Test VoltageMINMAX}
Unit: Volts
Return Value: {Test Voltage} in {NR3} format
Function: Set or query the test voltage for the LC Test.
{Voltage} 1.0 -650V
MIN 1.0V
MAX 650V
3.2.5.15 LCT:SOUR:CURR
Instruction: LCT:SOUR:CURR
Parameter: {Test CurrentMINMAX}
Unit: Milliamps
Return Value: {Test Current} in {NR3} format
Function: Set or query the test current for the LC Test.
{current} 0.5mA – 500mA
MIN 0.5mA
MAX 500mA (150mA for V>100V)
3.2.5.16 LCT:CONF:FUNC
Instruction: LCT:CONF:FUNC
Parameter: {SEQSTEP}
Return Value: {SEQSTEP}
Function: Set or query the configuration of the LC Test.
Description: SEQ LC Test is a Sequence Test
STEP LC Test is a Single Manual Test

Related product manuals