Page 64 of 76 Interface
3.2.5.17 LCT:CONF:SPE
Instruction: LCT:CONF:SPE
Parameter: {FASTMEDIUMSLOW}
Return Value: {FASTMEDIUMSLOW}
Function: Set or query the Measurement Speed.
Description: FAST 18 measurements/second
MEDIUM 14 measurements/second
SLOW 7 measurements/second
3.2.5.18 LCT:CONF:RANG
Instruction: LCT:CONF:RANG
Parameter: {<range>MINMAX}
Return Value: {<range>}
Function: Set or query the measurement range for the LC Test.
Description: 4 20mA
3 2mA
2 200uA
1 20uA
0 2uA
MIN 2uA
MAX 20mA
3.2.5.19 LCT:CONF:RANG:AUTO
Instruction: LCT:CONF:RANG:AUTO
Parameter: {OFFON01}
Return Value: {01}
Function: Set or query if the Auto Range function is OFF or ON.
Description: 0 Auto Range is OFF
1 Auto Range is ON
3.2.5.20 LCT:CONF:CHGT
Instruction: LCT:CONF:CHGT
Parameter: {<numeric value>MINMAX}
Return Value: {<numeric value>}
Function: Set or query the charge time for the LC Test.
Description: <numeric value> 0 – 999seconds
MIN 0seconds
MAX 999seconds

Related product manuals