3.2.5.21 LCT:CONF:DWEL
Instruction: LCT:CONF:DWEL
Parameter: {<numeric value>MINMAX}
Return Value: {<numeric value>}
Function: Set or query the dwell time for the LC Test.
Description: <numeric value> 0.2 – 999seconds
MIN 0.2seconds
MAX 999seconds
3.2.5.22 LCT:MEAS:STAT?
Instruction: LCT:MEAS:STAT?
Parameter: None
Return Value: {CHGTESTDCHG}
Function: Query the test status of the LC Test.
Description: CHG Instrument is in Charge mode
TEST Instrument is in Test mode
DCHG Instrument is in Discharge mode
3.2.5.23 LCT:MEAS:FETC?
Instruction: LCT:MEAS:FETC?
Parameter: None
Return Value: {01}, {ONPASSHIGHLOW}
Function: Query the test result of the LC Test.
Description: 0 Okay
1 Error
ON Instrument in Test mode
PASS Test Passed
HIGH Test Failed result above High Limit
LOW Test Failed – result below Low Limit
3.2.5.24 LCT:MEAS:IR?
Instruction: LCT:MEAS:IR?
Parameter: None
Return Value: {Measured value} in <NR3> format
Function: Query the IR measurement value.
Description: IR measurement value