Interface Page 67 of 76
3.2.5.29 WVT:CONF:TEND
Instruction: WVT:CONF:TEND
Parameter: {<numeric value>⏐MIN⏐MAX}
Unit: seconds
Return Value: {Measurement Time} in <NR3> format
Function: Set or query the measurement time for the WV Test.
Description: <numeric value> 30 – 600seconds
MIN 30seconds
MAX 600seconds
3.2.5.30 WVT:CONF:CHGTEND
Instruction: WVT:CONF:CHGTEND
Parameter: {<numeric value>⏐MIN⏐MAX}
Unit: seconds
Return Value: {Maximum Charge Time} in <NR3> format
Function: Set or query the maximum charge time for the WV Test.
Description: <numeric value> 5 – 600seconds
MIN 5seconds
MAX 600seconds
3.2.5.31 WVT:MEAS:STAT?
Instruction: WVT:MEAS:STAT?
Parameter: None
Return Value: {CHG⏐TEST⏐DCHG}
Function: Query the test status of the WV Test.
Description: CHG Instrument is in Charge mode
TEST Instrument is in Test mode
DCHG Instrument is in Discharge mode
3.2.5.32 WVT:MEAS:VT?
Instruction: WVT:MEAS:VT?
Parameter: None
Return Value: {Test Voltage} in <NR3> format
Function: Query the Test Voltage at the Termination of the WV Test.
Description: <numeric value> 1.0 – 650volts