Page 70 of 76 Interface
3.2.5.43 SYST:CONT
Instruction: SYST:CONT
Parameter: {<numeric value>}
Return Value: {Contrast} in <NR1> format
Function: Set or query the contrast of the display.
Description: <numeric value> 1-16
3.2.5.44 SYST:RANGED
Instruction: SYST:RANGED
Parameter: {<numeric value>MINMAX}
Return Value: {Dwell Time} in <NR3> format
Function: Set or query the range dwell time.
Description: <numeric value> 0 – 9.9seconds
MIN 0seconds
MAX 9.9seconds
3.2.5.45 SYST:AVER
Instruction: SYST:AVER
Parameter: {<numeric value>MINMAX}
Return Value: {Average} in <NR1> format
Function: Set or query the number of measurements made & averaged before result shown.
Description: <numeric value> 1-8
MIN 1
MAX 8
3.2.5.46 SYST:PRES
Instruction: SYST:PRES
Parameter: None
Return Value: None
Function: Set the instrument to initial default values.
3.2.5.47 SYST:ERR?
Instruction: SYST:ERR?
Parameter: None
Return Value: Error message
Function: Query if there are any system errors.
Description: <numeric value>, <string>