Interface Page 69 of 76
3.2.5.38 SYST:BEEP[:IMM]
Instruction: SYST:BEEP[:IMM]
Parameter: None
Return Value: None
Function: Set the beeper to sound immediately.
3.2.5.39 SYST:BEEP:STAT
Instruction: SYST:BEEP:STAT
Parameter: {OFF (0)ON, LOW (1)ON, HIGH (2)}
Return Value: {012}
Function: Set the loudness of the beeper.
Description: 0 Turn Beeper OFF
1 Set Beeper sound to LOW
2 Set Beeper sound to HIGH
3.2.5.40 SYST:ALAR
Instruction: SYST:ALAR
Parameter: {PULSCONT}
Return Value: {PULSCONT}
Function: Set the mode the alarm will sound in.
Description: PULS The alarm sound will pulse
CONT The alarm will continuously sound
3.2.5.41 SYST:LFR
Instruction: SYST:LFR
Parameter: {5060}
Unit: Hz
Return Value: {5060}
Function: Set or query the Line Frequency.
Description: 50Hz AC Power Line Source is 50Hz
60Hz AC Power Line Source is 60Hz
3.2.5.42 SYST:HAND
Instruction: SYST:HAND
Parameter: {CLEAHOLD}
Return Value: {CLEAHOLD}
Function: Set the Handler to clear result or hold result for each test
Description: CLEA Handler will clear result after each test
HOLD Handler will hold result after each test