EasyManua.ls Logo

RAUMEDIC NeuroSmart - Error Handling and Troubleshooting

RAUMEDIC NeuroSmart
61 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
RAUMED NeuroSmart Instruction Manual
Manufacturer: RAUMEDIC AG, Hermann-Staudinger-Strasse 2, 95233 Helmbrechts, Germany
zwo-541EN Rev 3 / 2022-05-03
Page 44 of 61
4 Error messages and troubleshooting
If an error occurs, the NeuroSmart shows an info message. Press the flashing speech bubble to
read the associated info message. The info messages with corresponding error code and troubleshoot-
ing recommendation are listed in Table 18.
Table 18: Physiological error message
Error
code
Error description
Troubleshooting
P01
The measured pressure P1/P2 is out-
side the set limits
P02
The measured temperature T1/T2 is
outside the set limits
P03
The calculated value CPP is outside
the set limits
P04
The measured pO2 value is outside
the set limits
Table 19: Technical sensor error message
Error
code
Error text
Troubleshooting
S01
The pressure sensor P1/P2 has been
removed. Please confirm. If you have
not removed the sensor, it is most
likely a sensor defect. Please check
this and replace the sensor if neces-
sary.
Confirm this with OK. If you have not removed the
sensor, it is probably a sensor defect. Please check
this and replace the sensor if necessary.
S03
The temperature sensor T1/T2 has
been removed. Please confirm. If you
have not removed the sensor, it is
most likely a sensor defect. Please
check this and replace the sensor if
necessary.
Confirm this with OK. If you have not removed the
sensor, it is probably a sensor defect. Please check
this and replace the sensor if necessary.
S04
The pO2 sensor P1/P2 has been re-
moved. Please confirm. If you have
not removed the sensor, it is most
likely a sensor defect. Please check
this and replace the sensor if neces-
sary.
Confirm this with OK. If you have not removed the
sensor, it is probably a sensor defect. Please check
this and replace the sensor if necessary.
S05
The connection with the NEU-
ROVENT P-tel was interrupted. The
NEUROVENT P-tel may be outside
the transmission range of the reader.
Check the Reader’s location and correct this if nec-
essary.
S06
The reader was disconnected from
the device
The Reader has been separated from the device.
If you have not separated the Reader, it is probably
a Reader defect.
S08
The multi parameter catheter ICP1/2
on P1/2 was disconnected from the
device
Confirm this with OK. If you have not removed the
Multi-parameter catheter, it is probably a sensor
defect. Please check this and replace the sensor if
necessary.
S09
The measured values are constant.
The catheter may be defective or im-
properly connected.
Confirm this with OK.
Check the pulsation of the pressure signals. If no
pulsating pressure signals are displayed even
though the catheter/transducer is connected to the
patient, it is probably a sensor defect.
This document is for information only

Table of Contents

Related product manuals