EasyManua.ls Logo

RAUMEDIC NeuroSmart - Setting Physiological Alarm Limits

RAUMEDIC NeuroSmart
61 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
RAUMED NeuroSmart Instruction Manual
Manufacturer: RAUMEDIC AG, Hermann-Staudinger-Strasse 2, 95233 Helmbrechts, Germany
zwo-541EN Rev 3 / 2022-05-03
Page 33 of 61
Settings options in the password-protected area
Set the time interval for AUDIO PAUSED and AUDIO OFF
Activate/deactivate the Reminder Signal [2]
System Reset: Reset to factory setting [3]
Change Password [4]
Export Logfiles [5].
The log is used for servicing and is encrypted. Please contact the service department. A USB stick must
be inserted to export the logfile. Click on the Export Logfile” button [5], which saves the logfile to the
USB stick.
Select the desired settings from <3> and confirm your entries <4>.
3.6.2 Setting physiological alarm limits
When the system is restarted, the default alarm limits are selected with the preset values shown in Table
16. The previous alarm limits or clinic-specific alarm limits can be adopted by selecting "Keep or Clinic"
in the alarm limits, see Figure 6, Alarm limit selection keys [2].
Please check the plausibility of the set alarm limits after the measurement has been started. See
Figure 21 and Figure 22.
Caution: ALARM LIMITS set to extreme values can make the ALARM SYSTEM unusable.
If the NeuroSmart is not continuously operated by clinical personnel, we recommend selecting the
“Default” alarm limits.
Table 16 Shows the default settings for the physiological alarms and their setting ranges.
Table 16: Default settings
Measured
variables
Measuring
position
Default settings
Setting range
Lower
alarm limit
Upper
alarm limit
Alarm ac-
tivated
Alarm limits *1
Lower upper
minimal
adjust-
ment dis-
tance
Print
ART
SYS
100
180
Yes
-40
400
5
Print
ART
MAP
80
140
Yes
-40
400
5
Print
ART
DIA
60
100
Yes
-40
400
5
Print
CVP
0
20
Yes
-40
400
1
Print
ICP1
0
30
Yes
-40
400
1
Print
ICP1
ICPA
0
20
No
0
400
1
Print
ICP2
0
30
Yes
-40
400
1
Print
ICP2
ICPA
0
20
No
0
400
1
Print
ICP1/2
CPP
50
100
Yes
0
400
1
Print
ICPT
0
30
Yes
-40
400
1
Print
pO2
10
80
Yes
0
200
1
Tempera-
ture
T1
35.8
37.5
Yes
15.0
45.0
0.1
Tempera-
ture
T2
35.8
37.5
Yes
15.0
45.0
0.1
*1 The alarm limits can be freely adjusted in the indicated range; the minimum distance between the
lower and upper alarm limit is called the minimum adjustment distance.
This document is for information only

Table of Contents

Related product manuals