EasyManua.ls Logo

RAUMEDIC NeuroSmart - Page 34

RAUMEDIC NeuroSmart
61 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
RAUMED NeuroSmart Instruction Manual
Manufacturer: RAUMEDIC AG, Hermann-Staudinger-Strasse 2, 95233 Helmbrechts, Germany
zwo-541EN Rev 3 / 2022-05-03
Page 34 of 61
The limits for the physiological alarms and their activation can be changed at any time. There are two
selection options.
Warning: If an alarm is deactivated, the alarm limits for this measuring position are not mon-
itored. This means that the acoustic and visual alarms are deactivated.
Initial approach: Adjust the alarm limits and alarm activation for individual active measuring po-
sitions.
<1> Press the alarm limit bar to access the alarm limit settings menu for individual measuring positions.
<2> The lower and upper alarm limit can be adjusted with the Plus/Minus buttons. <3> If you wish to
change the alarm activation status, you can do so with the “Alarm Active” button. <4> To make the
change, quit the menu with the Confirm button. See Figure 21.
[1] The default preset alarm limits can be reset with the “Reset to Default” button as per Table 16.
See Figure 21.
[2] If you do not want to make any change but wish to quit the menu, tap the “Cancel” button. See
Figure 21.
Figure 21: Adjust the alarm limits and alarm activation for the individual measuring positions.
This document is for information only

Table of Contents

Related product manuals