RAUMEDIC NeuroSmart - Environmental and Operating Conditions

RAUMEDIC NeuroSmart
61 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
RAUMED NeuroSmart Instruction Manual
Manufacturer: RAUMEDIC AG, Hermann-Staudinger-Strasse 2, 95233 Helmbrechts, Germany
zwo-541EN Rev 3 / 2022-05-03
Page 58 of 61
USB-A
USB type
USB 2.0 Standard
Connection
With a USB stick
Defibrillator protection
No
USB PC interface
2.0
Connection
With a PC or CSN Device Interface (Moberg)
Defibrillator protection
No
Ethernet interface
(Diagnosis interface)
Type:
Ethernet
Connection
With a PC for servicing
Defibrillator protection
No
8.2 Ambient conditions
8.2.1 Operating conditions
Temperature:
+ 10 °C ... + 40 °C
Air humidity:
between 15 and 75%
Air pressure:
between 620 and 1060 hPa
Light:
Protected from sunlight.
8.2.2 Storage conditions
Temperature:
- 10 °C ... + 40 °C
Air humidity:
between 15 and 75%
Air pressure:
between 620 and 1060 hPa
Light:
Protected from sunlight.
A longer storage over 35°C reduces the useful life of the built-in batteries. The device should always be
stored with fully charged batteries. At storage temperatures of more than 35°C, the battery must be
recharged four times a year, while at lower temperatures it only needs to be charged once a year.
8.2.3 Transport conditions
Temperature:
-25 °C ... + 60 °C
Air humidity:
Non-condensing between 15% and 93%
Air pressure:
620-1060 hPa
This document is for information only