RAUMED NeuroSmart Instruction Manual
Manufacturer: RAUMEDIC AG, Hermann-Staudinger-Strasse 2, 95233 Helmbrechts, Germany
zwo-541EN Rev 3 / 2022-05-03
Page 7 of 61
• With all securing and installation variants, ensure that the device cannot fall down or cause any
hazard in any other way. Route the cable safely to prevent tilting.
• Do not use the device in an explosive environment. The device must have at least 25 cm clearance
from any anaesthesia devices and gas-filled tubes.
• The accessories must be cleaned, disinfected or sterilised in accordance with the instruction man-
ual. If the accessories have a separate instruction manual, its instructions apply.
• Should the device be moved from a cold environment into a warm environment, it must only be used
once the temperature of the device and the accessories have reached room temperature and there
is no risk of condensation forming or any condensation formed has completely dried out.
• Do not allow any liquid to get into the device. However, if liquid does get in, firstly remove the adapter
from the wall outlet and give the device to Service for checking. Then carry out checks as per the
safety requirements.
• For installation of the system (with laptop or external system – patient monitor), the following instruc-
tions must always be strictly observed:
• Non-stationary universal portable socket outlets must not lie on the ground
• An additional non-stationary universal portable socket outlet or extension cable must not be
connected to the system
• Equipment that is not part of the system must not be connected to this
2.2 Instructions for cyber and data security
• The NeuroSmart is not intended to be operated with the Internet.
• The NeuroSmart is not intended for operation in public networks.
• Only connect the NeuroSmart to a USB stick or virus-free PC.
• Please delete all data before sending the NeuroSmart for routine service checks.
• Please delete all data at the end of the device’s service life if the NeuroSmart is to be perma-
nently shut down.
• Only persons/organisations authorised by RAUMEDIC may perform a software/firmware up-
date.
This document is for information only