device. The catheter can be used without using
pO2 measurement.
T26
There is a wrong NEUROVENT P-tel connected. If
you want to continue measurement with this NEU-
ROVENT P-tel you have to switch off the device
and start a new measurement after switching it on
again
T27
There is probably a hardware error in the NEU-
ROVENT P-tel
Contact Service
T28
There is an inconsistency in the data of the NEU-
ROVENT P-tel. A measuring process can not be
carried out. Please contact service
Contact Service
T29
The software version of the NEUROVENT P-tel is
not compatible with the NeuroSmart. Use another
NEUROVENT p-tel or contact service.
Contact Service
T30
There is an inconsistency in the data of the reader.
A measuring process can not be carried out. Use
another reader.
Contact Service
T31
The software version of the Reader is not compat-
ible with the NeuroSmart. A measuring process
can not be carried out. Use another reader or con-
tact the service staff for the software update.
Contact Service
T32
The version of the correction parameters in the
reader is not compatible with the measurement
system. A measuring process can not be carried
out. Use a different reader or contact the service
staff for the software update.
Contact Service
T33
No connection to the NEUROVENT P-tel possible.
Check the position of the reader antenna
Check positioning of the ring antenna
T34
P1/P2: The pO2 channel is already occupied The
NeuroSmart can only display one pO2 value
T35
P1/P2 self-test failed
Contact Service
T36
Rec readback error! Transfer to Rec has been dis-
abled.
Contact Service
T37
System time failure. Check Date and Time are cor-
rect.
Check date and time. Correct date
and time if necessary.
T50
Unexpected restart. Measurement is continued.
T51
Battery fault! Please contact service
Contact Service
T52
Battery defective! Please replace the battery.
Contact Service
T53
Loudspeaker faulty. Contact Service
Contact Service
T54
Piezo/buzzer faulty. Contact Service
Contact Service
T55
Internal device error. Perform a self-test or restart
the device if necessary. If the error occurs again,
please contact Service staff.
Contact Service
T56
Internal data error. Perform a self-test or restart the
device if necessary. If the error occurs again,
please contact Service staff.
Contact Service
T57
Internal CPU error. Perform a self-test or restart the
device if necessary. If the error occurs again,
please contact Service staff.
Contact Service
T58
Internal data error. Perform a self-test or restart the
device if necessary. If the error occurs again,
please contact Service staff
Contact Service
T59
System self-test failed. Perform a self-test or re-
start the device if necessary. If the error occurs
again, please contact Service staff
Contact Service