Introduction
5
SmartMonitor
No Apnea with Alarm
(False Alarm)
SmartMonitor 2
No Apnea with Alarm
(False Alarm)
Difference
(SmartMonitor 2 – SmartMonitor)
% Difference
97 85 -12 -12.3%
The false apnea rate for SmartMonitor is 97/(95.95) = 1.01.
The false apnea rate for SmartMonitor 2 is 85/95.95 = 0.89.
SmartMonitor Apnea
with No Alarm
(Missed Event)
SmartMonitor 2
Apnea with No Alarm
(Missed Event)
Difference
(SmartMonitor 2 – SmartMonitor)
% Difference
74 69 -5 -6.8%