Setting Alarm and Recording Limits
26
Menu Option LCD Display Description of Option or Instructions Menu Path
Time To Call
the Computer
WHEN TO
CALL
To enter the Time to call:
Press t or u until WHEN To CALL is
displayed.
Press Enter to activate.
Press t or u to browse through the
numbers and select the digit you need.
When the digit you want displays, press
Enter.
Enter the time in Military Time
.
When you have finished, press Enter.
Press t to proceed to the next menu or u
to move to the previous menu.
All Menus
System Menu
Call
Computer
when
memory is full
DIAL IF
MEMORY FULL
You may select from On or Off.
Once the Memory Full Alert is reached the
monitor will attempt to call at the next scheduled
calling time.
As an example, if the Dial If Memory Full is
turned ON and the call date and time is set to
3-1-01 at 4:00AM, but the memory reaches the
Memory Full Alert at 3:00PM on 2-1-01, then the
monitor will initiate a call at 4:00AM on 2-2-01.
All Menus
System Menu
Move Data to
Card
MOVE DATA
TO CARD?
(other possible
messages)
Insert Card –
Retry
Transferring
Data…
Data
Transferred
This menu is used to down load the recorded
data to the PCMCIA card. To move data to the
PCMCIA card:
Press the ENTER button. The word NO will
begin to blink.
To select YES press either arrow button.
Should a PCMCIA card not be inserted the
monitor will ask for the card.
Place card in slot located just below the
display.
Press the ENTER button.
The display will now show “Transferring
Data….”. Once the transfer is complete the
display will show “Data Transferred”.
All Menus
System Menu
Memory
Status
MEMORY
STATUS
X PERCENT
FULL
This is a display-only menu. It provides an
indication of the amount of memory in use by the
SmartMonitor 2. This display cannot be changed.
The Memory Status setting takes into account
the Patient Event Log, Equipment Event Log,
and Waveform Memory.
All Menus
System Menu
Clear
Memory
CLEAR
MEMORY?
Permits the memory to be erased. NOTE:
Erasing memory does not affect the alarm and /
or the record parameters.
The memory should be cleared before using
the SmartMonitor 2 on a new patient or after
the data has been downloaded and saved.
Once the memory is cleared the unit will
display the following:
“It Is Cleared”.
All Menus
System Menu