4.2.5 Access Code
To enter the Advanced Function and all function after the Advanced Function page a higher Access Code is
necessary. Access code for level 2 is 1802.
To enter this Access Code please follow below procedure:
On the Start Page press
In Main Menu scroll to Advanced Function, using and buttons.
Access Code page will pop-up if you are not already in Access Code 2
Enter Access Code using and buttons.
After entering the Access Code press the button.
The Advanced Function page is on the screen now.
4.2.6 Advanced Function
On the Advanced Function page you can choose between 5 options:
Calibration Menu, see chapter 5
Diagnostics Menu, see chapter 7
Liquid Select
Check Function
Output mA Range
4.2.7 Liquid Select
On the Liquid Select page you can choose 2 options:
Water measurement: only water can be measured, polynomials for water are standardly stored in
the analyzer.
Normal operation: a set of polynomials is loaded into the analyzer. This setting is specifically for
its application.