SDM Slurry Density Meter
47 www.rhosonics.com
CALIBRATION LCD SCREEN
Offset Menu page:
The offset shifts the indicated SGx1000 to the desired value, without influencing the sensitivity.
Enter S.G. x 1000 Offset using and buttons.
After entering the Offset press the button.
Span Menu page:
For some applications the sensitivity of the SDM is either too high or too low. When that is the case the
span has to be adjusted. The way of configuring the span factor is described in earlier in this chapter.
It is however not necessary to completely understand that explanation.
By changing the span you get the expected result of the calibration
Enter S.G. x 1000 Span using and buttons.
After entering the Span press the button.
5.1.3 Field calibration
Field calibration has no effect for the SDM.