RIGOL  Chapter 11 Waveform Recording 
11-16  MSO4000/DS4000 User’s Guide 
During the waveform analysis, you can press Cancel to stop the analysis. 
Pressing Start again can restart the analysis. 
Note: During the analysis process, the progress bar is displayed on the screen 
and the settings cannot be modified.  
 
5.  View the analysis results 
After the analysis finishes, the analysis results are displayed on the screen and 
the first error frame is located as shown in the figure below.  
 
 
 
 
1)  Waveform analysis result explanations 
  Error Frames: the total number of error frames discovered in this 
analysis. The number of error frames is related to the error frame 
threshold currently set.  
  Current Error: when the waveform analysis finishes, 1 (error frame is 
discovered) or 0 (no error frame is discovered) is displayed by default. 
When viewing the error frames, the order number of the error frame 
currently viewed in all the error frames is displayed.  
  CurFrame Diff: related to the analysis mode currently selected.  
 
Trace Mode 
The “CurFrame Diff” (namely the relative difference) is the normalized 
value of the difference between the current frame and template. Please 
refer to “Trace Mode”.  
 
Pass/Fail Test Mode 
The “CurFrame Diff” is 100% (the frame is correct) or 0% (the frame is 
wrong). Please refer to “Pass/Fail Test Mode”.  
 
Note: There are only two “CurFrame Diff” values (100% and 0%) 
Waveform Analysis Settings and Results 
www.calcert.com sales@calcert.com1.800.544.2843
0
5
10
15
20
25
30