Chapter 2 Vertical System Setting    RIGOL 
MSO4000/DS4000 User’s Guide  2-3 
To Adjust the Vertical Scale 
 
The vertical scale refers to the voltage value per grid on the vertical direction on the 
screen and is usually expressed as V/div. The adjustable range of the vertical scale is 
related to the current probe ratio and input impedance. By default, the probe ratio is 
1X, the input impedance is 1 MΩ and the adjustable range of the vertical scale is from 
1 mV/div to 5 V/div.  
 
Press CH1 and its backlight is illuminated; then, press Volts/Div to switch the 
vertical scale adjustment mode of CH1 between “Coarse” and “Fine” (the default is 
“Coarse”). You can also press VERTICAL 
  SCALE corresponding to CH1 to 
quickly switch the current adjustment mode. 
 
  When “Coarse” is selected, turning VERTICAL    SCALE will adjust the 
vertical scale within the adjustable range at 1-2-5 step sequence. Turn the knob 
clockwise to reduce the vertical scale and counterclockwise to increase the 
vertical scale. 
  When “Fine” is selected, turning VERTICAL    SCALE will further adjust the 
vertical scale within the adjustable range at a smaller step to improve the 
vertical resolution. Turn the knob clockwise to reduce the vertical scale and 
counterclockwise to increase the vertical scale.  
If the amplitude of the input waveform is a little bit greater than the full scale of 
the current scale and the amplitude would be a little bit lower if the next scale is 
used, fine adjustment can be used to improve the waveform display amplitude 
to view signal details. 
 
When modifying the vertical scale, the waveform of the corresponding channel will 
be expanded or compressed vertically relative to the vertical reference currently 
selected (refer to “To Specify the Vertical Reference”); the vertical scale 
information (as shown in the figure below) in the channel status label at the bottom 
of the screen will change in real time. 
 
 
 
www.calcert.com sales@calcert.com1.800.544.2843
0
5
10
15
20
25
30