RIGOL  Chapter 1 Quick Start 
1-22  MSO4000/DS4000 User’s Guide 
TRIGGER 
 
 
  MODE: press this key to switch the trigger mode to “Auto”, 
“Normal” or “Single” and the corresponding status backlight 
of the current trigger mode would be illuminated. 
 
  TRIGGER 
  LEVEL: trigger level knob. 
Turn the knob to modify the trigger level. During the 
modification, the trigger level line moves up and down; the 
trigger level message (as shown in the figure below) pops up 
at the lower-left side of the screen in real-time. Pressing 
down the knob can quickly reset the trigger level to 0. 
 
 
  MENU: press this key to open the trigger operation menu. 
This oscilloscope provides various trigger types. 
 
  FORCE: in Normal and Single trigger modes, press this key 
to generate a trigger signal forcibly. 
 
  50%: press this key to set the trigger level to the vertical 
midpoint of the trigger signal amplitude. 
 
CLEAR 
 
 
Press this key to clear all the waveforms on the screen. If the 
oscilloscope is in “RUN” state, new waveforms will be 
displayed. 
 
 
 
 
 
 
 
 
 