? CRLF Command not understood
0 CRLF Command correctly operated
1 CRLF Scale in motion
2 CRLF Range out. For example, zero setting out of range
3 CRLF System error
4 CRLF Device not ready. For example, printer not ready
Example
Command Answer Description
S01; Select unit 1
MSV?; 00400.0 CRLF Query current weight
TAR; 0 CRLF Tare
MSV?; 00000.0 CRLF Query current weight reading
MSV?1; 00400.0 CRLF Query gross weight
TAR?; 4000 CRLF Query current tare weight
21.3.49 TAS: Gross / Net
Select Gross or Net weight display.
General
No. of parameters 1
Save changes At input
Increment Trade Counter No
Parameter Details
Parameter Description Range Default
1 Gross or Net 0 = Net
1 = Gross
-
Table of possible answers:
Table of possible replies:
? CRLF Command not understood
0 CRLF Command correctly operated
1 CRLF Scale in motion
2 CRLF Range out. For example, zero setting out of range
3 CRLF System error
4 CRLF Device not ready. For example, printer not ready
Example
Command Answer Description
S01; Select unit 1
MSV?; 00200.0 CRLF Query current weight
TAS?; 0 CRLF Unit is in net mode
TAS1; 0 CRLF Switch to Gross weight
MSV?; 00400.0 CRLF Query current weight
TAS?; 1 CRLF Unit is in gross mode
21.3.50 TAV: Tare Value
Set a numeric tare value directly.
General
155 C500-600-2.0.3