5. Press the cancel key to exit the direct zero calibration and return to the menu.
6. Press the tare key to show the CAL:DIR.SPN setting (displays DIR.SPN).
7. Press the OK key. The display will show the current weight.
8. Press the OK key to enter the mV/V value of span at fullscale. Use the left/right/up/down keys to edit
the mV/V value.
9. Press the OK key to enter the calibration. The display will perform the calibration and then display the
weight to allow the calibration to be checked.
10. Press the cancel key to exit span calibration and return to the menu.
7.7 CAL:FAC.CAL Factory Calibration
The calibration can be reset to the factory calibration using the CAL:FAC.CAL menu item.