Roche Diagnostics
B-16 Operatorโ€™s Manual ยท Version 3.1
5 Daily operation Cedex Bio System
Preparing the system
a To check the water container
1
Check the water button.
2
Refill the water, if required.
o Remove the tubing adapter from the white water container and place it on a
clean surface.
o Refill the bottle with purified water.
o Insert the tubing adapter. Press down firmly.
3
On the screen that shows the container status, press the water button to confirm
that you have refilled the water.
4
Roche recommend to empty the waste container whenever you refill the water
container.
e
See To check the waste container on page B-17.
There is no water left. (A fluid sensor detected that there is no water coming
from the water container.)
You need to refill the water now. No new tests can be processed, started tests
may have to be restarted after refill.
The water container was last filled up more than one day ago.
You should check the water level and refill if required.
No action is required.
The system does not monitor the filling level. It is up to the operator to fill up the water
container regularly.
WARNING
Danger of poor measurement quality due to inadequate water quality
Inadequate water quality may lead to incorrect results. Always use purified water of the
quality specified in section Technical specifications.
When refilling the water without using the Prepare wizard:
Choose Overview > > .

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche Cedex Bio System and is the answer not in the manual?

Roche Cedex Bio System Specifications

General IconGeneral
BrandRoche
ModelCedex Bio System
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals