EasyManuals Logo

Roche Cedex Bio System User Manual

Roche Cedex Bio System
496 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #147 background imageLoading...
Page #147 background image
Roche Diagnostics
Operatorโ€™s Manual ยท Version 3.1 B-17
Cedex Bio System 5 Daily operation
Preparing the system
a To check the waste container
1
Check the waste button.
2
Empty the waste container, if required.
o Have the spare waste container ready.
The system periodically performs wash actions. Therefore, an external waste
container must be connected at all times.
o Remove the tubing adapter from the yellow waste container and insert it in the
spare container.
o Remove the bottle from the rack and place it on a firm and even surface.
o Place the spare container on the rack.
o Check that the tubing adapter is properly inserted.
o Empty the removed waste container. Treat the fluid as biohazardous waste.
o Rinse the container with water and leave it to dry.
3
On the screen that shows the container status, press the waste button to confirm
that you have emptied the container.
The waste container was last emptied more than one day ago.
You should check the waste level and empty if required.
No action is required.
The system does not monitor the filling level. It is up to the operator to empty the waste
container regularly.
When emptying the waste without using the Prepare wizard:
Choose Overview > > .

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche Cedex Bio System and is the answer not in the manual?

Roche Cedex Bio System Specifications

General IconGeneral
BrandRoche
ModelCedex Bio System
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals