5 Daily operation Cedex Bio System
Preparing the system
7
Open the main cover and remove the bottle.
8
Press to confirm the removal.
The system does not check whether you have actually removed the bottle.
9
Do one of the following:
The screen with the reagent set buttons is displayed again. The button for the
removed set is no longer present. If you did not remove all bottles of the set, the
reagent set would be disabled, its button would be red.
Adding a reagent set
10
Open the main cover if required.
11
Press .
A screen is displayed, asking you to scan the bottle barcode.
12
Scan the barcode on the bottle.
A screen is displayed, asking you to place the bottle on the reagent disk.
If Do this
There is another bottle
belonging to the set:
The system moves this bottle to the reagent port. A screen
is displayed, asking you to remove the bottle.
1. Remove the bottle.
2. Press to confirm the removal.
3. Start adding the new reagent set.
All bottles of the set are
removed:
1. Start adding the new reagent set.
Before you add a set, check the number to the left of the button, it indicates the num-
ber of free positions on the disk. If this number is smaller than the number of bottles in
the set you need to add, remove a set first. For the number of bottles in a set refer to
the respective package insert.
When scanning, bear this in mind:
o Scanning its barcode is the only way of identifying a reagent, diluent, and cleaner
bottle.
o If there are not enough free positions for all bottles of the set, a message will inform
you.
o You cannot re-insert empty bottles that were previously removed from the system.
If you scan such bottle, a message will inform you of this fact.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche Cedex Bio System and is the answer not in the manual?

Roche Cedex Bio System Specifications

General IconGeneral
BrandRoche
ModelCedex Bio System
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals