Roche Diagnostics
Operatorโ€™s Manual ยท Version 3.1 B-27
Cedex Bio System 5 Daily operation
Preparing the system
13
Remove the cap from the bottle and place it in the reagent bay.
14
Press to confirm the insertion.
If you confirm without placing the bottle, the system assumes that the bottle is
placed.
If you press after placing the bottle, the position is deemed empty.
When the first reagent bottle is loaded, the reagent set is defined. From this
moment on, the reagents are handled as part of the set. You no longer handle
them as individual reagents.
15
Do one of the following:
16
On the screen with the reagent set buttons, press the button for the set you just
inserted.
A screen is displayed that shows details of the set.
The status description points to the action you need to take. If you just inserted a
reagent set, you may have to perform the initial calibration and QC.
Time limit for placing reagent bottles
o The system assumes that you place the bottle that you just scanned.
o You need to place the reagent bottle on the reagent disk within 15 seconds of
scanning the reagent barcode. Failing to confirm the placement within 15 seconds
cancels the current identification process. You will be asked to scan the bottle
again.
WARNING
Incorrect results due to not placing the identified reagent
The system assumes that the operator places the reagent that was just identified. Fail-
ing to do so might lead to wrong results.
If Do this
There is another bottle
belonging to the set:
A screen is displayed, asking you to scan the bottle.
1. Scan the bottle
2. Insert the bottle.
3. Press to confirm the insertion.
All bottles of the set are
inserted:
1. Close the main cover.
On the screen with the reagent set buttons, the button
for the new set is now present.
A Status description
Figure B-7
A

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche Cedex Bio System and is the answer not in the manual?

Roche Cedex Bio System Specifications

General IconGeneral
BrandRoche
ModelCedex Bio System
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals