Roche Diagnostics
Operatorโs Manual ยท Version 3.1 B-65
Cedex Bio System 5 Daily operation
Performing calibrations
a To validate the calibration results
1
Choose Workplace > Calibrations
A screen is displayed that lists all tests and their currently valid calibration.
Interpreting the screen
e
For details on calibration types, see Calibration type on page A-26.
2
Select the result.
A Test ID.
B Calibration use.
C Calibration type.
D Status indication.
E Press to validate the result.
F Press to define a new calibration order.
G Press to delete the selected calibration
and its results.
H Press to display all or only the current
results.
Figure B-22
Item Value Interpretation
U (Use)
CU Current.
The calibration results of this set are currently used.
SB Standby.
The calibration results of this set are not currently used.
(An identical set is currently in use or the set was removed
and is not empty.)
OB Obsolete.
The calibration results of this set are no longer used. (The
set was empty and removed or it was removed more than
30 days ago).
T (Type)
L Lot calibration (first calibration of a lot, also applies to sub-
sequent sets of the lot).
S Set calibration (applies to calibrated set only).
Status
The date indicates when the results were accepted. If flags
were generated for the result, the flag with the highest pri-
ority is displayed. In all other cases, the order status is dis-
played.