EasyManuals Logo

Roche Cedex Bio System User Manual

Roche Cedex Bio System
496 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #248 background imageLoading...
Page #248 background image
Roche Diagnostics
B-118 Operatorโ€™s Manual ยท Version 3.1
6 Special operations Cedex Bio System
Preparing a new disk
Loading reagent sets
9
Press .
A screen is displayed, asking you to scan the bottle barcode.
10
Scan the barcode on the bottle.
11
Remove the cap from the bottle and place it in the reagent bay.
12
Press to confirm the insertion.
If you confirm without placing the bottle, the system assumes that the bottle is
placed.
If you press after placing the bottle, the position is deemed empty.
When the first reagent bottle of a set is loaded, the reagent set is defined. From
this moment on, the reagents are handled as part of the set. You no longer handle
them as individual reagents.
13
Do one of the following:
14
Press .
15
On the disk overview screen, press to load the next reagent. Proceed as
described in steps 10 through 15.
If Do this
There is another bottle
belonging to the set:
A screen is displayed, asking you to scan the bottle.
1. Scan the bottle
2. Insert the bottle.
3. Press to confirm the insertion.
All bottles of the set are
inserted:
1. Close the main cover.
On the screen with the reagent set buttons, the button
for the new set is now present.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche Cedex Bio System and is the answer not in the manual?

Roche Cedex Bio System Specifications

General IconGeneral
BrandRoche
ModelCedex Bio System
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals