Roche Diagnostics
Operatorโ€™s Manual ยท Version 3.1 D-11
Cedex Bio System 9 Messages and alarms
List of alarm messages
0020 This reagent set has more bottles than there is space
available on the disk.
Load another disk or remove unused reagents.
0021 Disk space problem.
0022 You are about to delete results that have been printed or
sent to host!
0023 Enter the A1c part.
0024 Now enter the A1c part.
5001 Air detected in the fluid system. Water container or
cleaner bottle is empty.
5002 [%d] Air detected in the fluid system. Refill the water container. Check the filling level of the
cleaner bottle. Replace the cleaner bottle if required. Per-
form <Prime Fluid System>. If the error persists call ser-
vice.
5003 [%s] An internal software error occurred. A parameter is
not valid.
Restart the system. If the error persists call service.
5004 [%s] Error syringe motor. The syringe motor could not be
started.
Restart the system. If the error persists call service.
5010 [%s] Initialization of syringe failed. Switch off instrument. Remove right side panel. Check for
syringe mechanism jamming or blockage. Ensure correct
syringe installation. Restart the system. If the error per-
sists call service.
5013 [%s] Error mixer motor. The mixer motor could not be
started.
Restart the system. If the error persists call service.
5019 [%s] Initialization of the mixer motor failed. Switch off the instrument. Remove the transfer head
cover. Check if the mixing mechanism is jamming or
blocked. Restart the system. If the error persists call ser-
vice.
5022 [%s] Error temperature sensor analyzer heating system.
The sensor does not react anymore.
Restart the system. If the error persists call service.
5024 [%s] Error temperature sensor analyzer cooling system.
The sensor does not react anymore.
Restart the system. If the error persists call service.
5026 [%s] Error temperature sensor cooling system. The heat
sink temperature sensor does not react anymore.
Restart the system. If the error persists call service.
5028 [%s] Error temperature sensor instrument. The ambient
temperature sensor does not react anymore.
Restart the system. If the error persists call service.
5030 [%s] Error wash pump P1. The wash pump could not be
started.
Restart the system. If the error persists call service.
5032 [%s] Error waste pump P2. The waste pump could not be
started.
Restart the system. If the error persists call service.
5034 [%s] Error wash valve V1. The valve does not react
anymore.
Restart the system. If the error persists call service.
5036 [%s] Error cleaner valve V2. The valve does not react
anymore.
Restart the system. If the error persists call service.
5038 [%s] Error analyzer heating system. The heating system
does not react anymore.
Restart the system. If the error persists call service.
5040 [%s] Error cooling system. One of the Peltier elements
does not react anymore.
Restart the system. If the error persists call service.
ID Message Comment
Table D-1 List of alarm messages (Continued)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche Cedex Bio System and is the answer not in the manual?

Roche Cedex Bio System Specifications

General IconGeneral
BrandRoche
ModelCedex Bio System
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals