Roche Diagnostics
D-12 Operatorโs Manual ยท Version 3.1
9 Messages and alarms Cedex Bio System
List of alarm messages
5044 [%s] Error main board fan. The fan could not be started or
the fan speed is not correct.
Restart the system. If the error persists call service.
5047 [%s] Error cooling fan. The fan could not be started. Restart the system. If the error persists call service.
5049 [%s] Error sample area fan. Switch off the instrument. Remove the right side panel and check if the fan is jam-
ming or blocked. If possible eliminate the problem.
Restart the system. If the error persists call service.
5052 [%s] Error cooling system. The cooling heat sink
overheated.
Ensure unrestricted cooling air circulation. Check the
cooling fan and make sure the external vent is not
blocked. Restart the system. If the error persists call ser-
vice.
5054 [%s] The analyzer heating system overheated. Switch off the instrument and wait 15 minutes to let cool
down the analyzer. Restart the system. If the error persists
call service.
5055 [%s] The analyzer temperature is outside the range. Restart the system. If the error persists call service.
5056 [%s] The calibration of the transfer head fluid sensor
failed.
Ensure that the sample probe tube is correctly installed.
Ensure that the syringe is correctly installed. Restart the
system. If the error persists call service.
5058 [%s] Error fluid sensor transfer head. The sensor does not
react anymore.
Restart the system. If the problem persists, call service.
5060 [%s] A not specified hardware error (%d) was reported by
the Multi-Slave control.
Restart the system. If the error recurs export the log files
with the system remaining in this error condition. If the
error persists call service.
5061 [%d] Air detected in the fluid system. Refill the water container. Check the filling level of the
cleaner bottle. Replace the cleaner bottle if required. Per-
form <Prime Fluid System>. If the error persists call ser-
vice.
5062 The degasser fluid sensor is not initialized. Perform <Initialize Degasser Fluid Sensor>. Follow the
instructions provided on the display. If the error persists
call service.
5063 [%s] Air bleed valve error. The air bleed valve does not
react anymore.
Restart the system. If the error persists call service.
5064 [%s] Error fluid sensor degasser. The sensor does not
react anymore.
Restart the system. If the error persists call service.
6001 [%s] Photometer error. Signal is invalid or unstable. Switch off the instrument. Remove and clean the rotor. In
case of a recurrence perform <Replace Photometer
Lamp> and <Abs Air/Water Calibration>. If the error
persists call service.
6002 [%s] Photometer error. The photometer has a wrong
status because initialization has not been successfully
performed yet.
Restart the system. If the error persists call service.
6018 [%s] A not specified photometer hardware error (%d) was
reported by the photometer.
Restart the system. If the error recurs export the log files
with the system remaining in this error condition. If the
error persists call service.
6019 [%s] Photometer error. The photometer signals were
outside the range while performing a photometer
initialization or a photometer measurement.
Restart the system. If the error persists call service.
ID Message Comment
Table D-1 List of alarm messages (Continued)