D-18 Operatorโs Manual ยท Version 3.1
9 Messages and alarms Cedex Bio System
List of alarm messages
7125 [%s] Motion control error in rotor motor. Restart the system to correct the problem. If the error per-
sists call service.
7126 [%s] Motion control error X-transfer motor. Restart the system to correct the problem. If the error per-
sists call service.
7127 [%s] Motion control error Z-transfer motor. Restart the system. If the error persists call service.
7128 [%s] Initialization error X-transfer. Switch off the instrument. Ensure that all covers are
closed. Check if the X-transfer is blocked or jamming.
Restart the system. If the error persists call service.
7129 [%s] Initialization error Z-transfer. Switch off the instrument. Check if the Z-transfer is
blocked or jamming. Check the seating of the transfer
head cover. Restart the system. If the error persists call
service.
7130 [%s] Rotor position error. Switch off the instrument and check if the rotor is blocked
or jamming. If the error recurs call service.
7131 [%s] Position error X-transfer. Switch off the instrument and check if the X-transfer is
blocked or jamming. If the error recurs call service.
7132 [%s] Initialization error Z-transfer. Switch off the instrument. Check if the Z-transfer is
blocked or jamming. Check the seating of the transfer
head cover. Restart the system. If the error persists call
service.
7133 [%s] Rotor motor overheated. Switch off the instrument. Check if the rotor is blocked or
jamming. Restart the system. If the error persists call ser-
vice.
7134 [%s] Motor X-transfer overheated. Switch off the instrument. Check if the X-transfer is
blocked or jamming. Ensure that the X-transfer can be
moved smoothly. Restart the system. If the error persists
call service.
7135 [%s] Motor Z-transfer overheated. Switch off the instrument. Check if the Z-transfer is
blocked or jamming. Ensure that the Z-transfer can be
moved smoothly. Restart the system. If the error persists
call service.
7136 [%s] Tube bottom detector is blocked or jamming. Switch off the instrument. Remove the transfer head
cover. Check the tube bottom detector mechanism. Install
the transfer head cover. Restart the system. If the error
persists call service.
7137 [%s] Na sample measurement unstable. Check the expiry date of the Ref and Na electrodes.
Replace expired electrodes. Check the ISE tubing for
obstructions and leaks. Perform <Electrode Service>. If
the error persists call service.
7138 [%s] Cl sample measurement unstable. Check the expiry date of the Ref and Cl electrodes.
Replace expired electrodes. Check the ISE tubing for
obstructions and leaks. Perform <Electrode Service>. If
the error persists call service.
7139 [%s] K sample measurement unstable. Check the expiry date of the Ref and K electrodes. Replace
expired electrodes. Check the ISE tubing for obstructions
and leaks. Perform <Activate Electrodes>. If the error per-
sists call service.
7140 [%s] Na calibrator measurement unstable. Perform <Electrode Service> and <Prime ISE Reference
and Calib.>. Check ISE Cal and Ref flow. Check the ISE
tubing for obstructions. If the error persists call service.