EasyManua.ls Logo

Roche Cedex Bio System - Page 423

Roche Cedex Bio System
496 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Roche Diagnostics
Operator’s Manual · Version 3.1 E-33
Cedex Bio System 13 ISE operation
Replacing electrodes
a To replace an electrode
1
Check that the system is in Standby status.
2
Choose Overview > .
The ISE status is displayed.
3
Press the button of the electrode you want to replace.
Details of the electrode status are displayed.
4
Press to start the exchange wizard.
The system rinses the electrodes with ISE Calibrator indirect/urine to remove any
sample residue and then drains all electrodes.
You are asked to remove the electrode. (Make sure to remove the electrode that is
marked on the screen.)
5
Open the ISE cover.
6
Open the electrode block lid.
7
Release the tension lever (A).
Turning the tension lever releases the electrodes. You may have to push the
anchoring piece towards the tension lever to create enough space for removing
the electrodes.
A Electrode buttons
Figure E-5
A Tension lever
Figure E-6
A
A

Table of Contents

Related product manuals