Command Reference
R&S
®
CMW500
202User Manual 1173.9463.02 ─ 02
the tables refer to RF Path Settings (Generators) and Connection Control (Measure-
ments).
Reset / Preset of base settings
Please note that a reset or preset of the base settings deactivates all correction tables.
A reset or preset of a sub-instrument or application does not affect the correction tables.
Commands deactivating correction tables:
●
SYSTem:RESet:BASE, SYSTem:PRESet:BASE
●
SYSTem:RESet:ALL, SYSTem:PRESet:ALL
Commands without impact on correction tables:
●
*RST
●
SYSTem:RESet, SYSTem:PRESet
CONFigure:BASE:FDCorrection:CTABle:CREate..............................................................202
CONFigure:BASE:FDCorrection:CTABle:ADD..................................................................203
CONFigure:BASE:FDCorrection:CTABle:DELete..............................................................203
CONFigure:BASE:FDCorrection:CTABle:CATalog.............................................................204
CONFigure:BASE:FDCorrection:CTABle:DETails..............................................................204
CONFigure:BASE:FDCorrection:CTABle:LENGth..............................................................204
CONFigure:FDCorrection:ACTivate.................................................................................205
CONFigure:FDCorrection:DEACtivate .............................................................................205
CONFigure:FDCorrection:USAGe....................................................................................206
CONFigure:BASE:FDCorrection:CTABle:CREate '<TableName>', <Frequency_0>,
<Attenuation_0>[, ..., <Frequency_n>, <Attenuation_n>]
Creates a new correction table for frequency-dependent attenuation and stores it on the
hard disk. If a table with the given name already exists, it is overwritten.
The parameter pairs <Frequency>, <Attenuation> are used to fill the table. At least one
parameter pair has to be entered. A command with an incomplete pair (e.g. <Frequency>
without <Attenuation>) is ignored completely. To add entries to an existing table see
CONFigure:BASE:FDCorrection:CTABle:ADD.
You can enter parameter pairs in any order. The table entries (pairs) are automatically
sorted from lowest to highest frequency.
Setting parameters:
'<TableName>' String parameter used to identify the table by other commands and
to store the table on the hard disk. The string must comply to Win-
dows™ file name conventions, see MMEMory Commands.
<Frequency...> Range: 100E+6 Hz to 6E+9 Hz
Increment: 0.1 Hz
Default unit: Hz
<Attenuation...> Range: -50 dB to 90 dB
Increment: 0.01 dB
Default unit: dB
Instrument-Control Commands