Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
227User Manual 1173.9240.02 ─ 17
Example: 
CORR:REC
Usage:  Event
Manual operation: See "Recall" on page 79
[SENSe:]CORRection[:STATe] <State>
This command turns correction of measurement results (normalization) on and off.
The command is available after you have created a reference trace for the selected
measurement type with [SENSe:]CORRection:COLLect[:ACQuire] on page 225.
This command is only available if external generator control is active (see SOURce:
EXTernal[:STATe] on page 222).
Parameters:
<State> ON | OFF
*RST:  OFF
Example: 
CORR ON
Activates normalization.
Usage:  SCPI confirmed
Manual operation: See "Source Calibration Normalize" on page 79
[SENSe:]CORRection:TRANsducer:GENerator <Name>
This command uses the normalized measurement data to generate a transducer factor
with up to 1001 points. The trace data is converted to a transducer with unit dB and
stored in a file with the specified name and the suffix .trd under
c:\r_s\instr\trd. The frequency points are allocated in equidistant steps between
start and stop frequency.
The generated transducer factor can be further adapted using the commands descri-
bed in the "Remote Commands > Configuring the R&S FSW > Working with Transduc-
ers" section in the R&S FSW User Manual.
Parameters:
<Name> '<name>
Example: 
CORR:TRAN:GEN 'SMU01'
Creates the transducer file C:\r_s\instr\trd\SMU01.trd.
Usage:  SCPI confirmed
Manual operation: See "Save As Trd Factor" on page 80
Programming Example for External Generator Control
The following example demonstrates how to work with an external generator in a
remote environment.
Configuring the Measurement