Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
277User Manual 1173.9240.02 ─ 17
11.5 Capturing Data and Performing Sweeps
MSRA/MSRT operating mode
Note that in MSRA/MSRT operating mode, capturing data is only possible for the
MSRA/MSRT Master channel. In Analog Demodulation application channels, the
sweep configuration commands define the analysis interval. Be sure to select the cor-
rect measurement channel before using these commands.
ABORt.......................................................................................................................... 277
INITiate:CONMeas.........................................................................................................278
INITiate:CONTinuous..................................................................................................... 278
INITiate[:IMMediate]....................................................................................................... 279
INITiate:REFResh..........................................................................................................279
INITiate:SEQuencer:ABORt............................................................................................ 280
INITiate:SEQuencer:IMMediate....................................................................................... 280
INITiate:SEQuencer:MODE.............................................................................................280
INITiate:SEQuencer:REFResh[:ALL]................................................................................281
SYSTem:SEQuencer......................................................................................................281
ABORt
This command aborts a current measurement and resets the trigger system.
To prevent overlapping execution of the subsequent command before the measure-
ment has been aborted successfully, use the *OPC? or *WAI command after ABOR and
before the next command.
For details see the "Remote Basics" chapter in the R&S FSW User Manual.
To abort a sequence of measurements by the Sequencer, use the INITiate:
SEQuencer:ABORt on page 280 command.
Note on blocked remote control programs:
If a sequential command cannot be completed, for example because a triggered sweep
never receives a trigger, the remote control program will never finish and the remote
channel to the R&S FSW is blocked for further commands. In this case, you must inter-
rupt processing on the remote channel first in order to abort the measurement.
To do so, send a "Device Clear" command from the control instrument to the
R&S FSW on a parallel channel to clear all currently active remote channels. Depend-
ing on the used interface and protocol, send the following commands:
●
Visa: viClear()
●
GPIB: ibclr()
●
RSIB: RSDLLibclr()
Now you can send the ABORt command on the remote channel performing the mea-
surement.
Capturing Data and Performing Sweeps