EasyManua.ls Logo

Rohde & Schwarz FSW Series - INITiate[:IMMediate]; INITiate:REFResh

Rohde & Schwarz FSW Series
371 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
279User Manual 1173.9240.02 ─ 17
Example:
INIT:CONT OFF
Switches the sweep mode to single sweep .
INIT:CONT ON
Switches the sweep mode to continuous sweep .
Manual operation: See "Continuous Sweep/RUN CONT" on page 104
INITiate[:IMMediate]
This command starts a (single) new measurement.
With sweep count or average count > 0, this means a restart of the corresponding
number of measurements. With trace mode MAXHold, MINHold and AVERage, the
previous results are reset on restarting the measurement.
You can synchronize to the end of the measurement with *OPC, *OPC? or *WAI.
For details on synchronization see the "Remote Basics" chapter in the R&S FSW User
Manual.
Manual operation:
See "Single Sweep/ RUN SINGLE" on page 104
INITiate:REFResh
This function is only available if the Sequencer is deactivated (SYSTem:SEQuencer
SYST:SEQ:OFF) and only for applications in MSRA/MSRT mode, not the MSRA/
MSRT Master.
The data in the capture buffer is re-evaluated by the currently active application only.
The results for any other applications remain unchanged.
The application channel must be selected before this command can be executed (see
INSTrument[:SELect] on page 179).
Example:
SYST:SEQ:OFF
Deactivates the scheduler
INIT:CONT OFF
Switches to single sweep mode.
INIT;*WAI
Starts a new data measurement and waits for the end of the
sweep.
INST:SEL 'IQ ANALYZER'
Selects the IQ Analyzer channel.
INIT:REFR
Refreshes the display for the I/Q Analyzer channel.
Usage: Event
Manual operation: See "Refresh" on page 105
Capturing Data and Performing Sweeps

Table of Contents

Other manuals for Rohde & Schwarz FSW Series

Related product manuals