Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
282User Manual 1173.9240.02 ─ 17
Parameters:
<State> ON | OFF | 0 | 1
ON | 1
The Sequencer is activated and a sequential measurement is
started immediately.
OFF | 0
The Sequencer is deactivated. Any running sequential measure-
ments are stopped. Further Sequencer commands (INIT:
SEQ...) are not available.
*RST: 0
Example:
SYST:SEQ ON
Activates the Sequencer.
INIT:SEQ:MODE SING
Sets single Sequencer mode so each active measurement will
be performed once.
INIT:SEQ:IMM
Starts the sequential measurements.
SYST:SEQ OFF
11.6 Configuring the Result Display
The following remote commands are required to configure the screen display in a
remote environment.
General Window Commands................................................................................ 282
Working with Windows in the Display................................................................... 283
11.6.1 General Window Commands
The following commands are required to configure general window layout, independent
of the application.
Note that the suffix <n> always refers to the window in the currently selected measure-
ment channel (see INSTrument[:SELect] on page 179).
DISPlay:FORMat........................................................................................................... 282
DISPlay[:WINDow<n>]:SIZE........................................................................................... 283
DISPlay:FORMat <Format>
This command determines which tab is displayed.
Configuring the Result Display